PhoSim Tutorial 9: How to Generate Calibration Data
Calibration data can be generated in a similar manner as with a real telescope. Basically, the commands have to be given to the
A bias frame can be generated with the catalog:
phosim examples/bias
The catalog simply has these two commands to set the exposure time to 0 and to make the minimum number of sources to be 0. This is necessary because PhoSim will not generate chips with no sources on them by default.
vistime 0.0
minsource 0
The electron image will be empty, and the amplifier images will have the appropriate bias level and read noise.
Similarly, a dark can be generated by closing the dome and setting the exposure time to a reasonably long value. An example of this is the following command.
phosim examples/dark
The example has the following commands that set the exposure time to 5 minutes, have the minimum source as 0, and close the dome with "telconfig 1".
vistime 300.0
minsource 0
telconfig 1
The electron image will contain dark current (if significant in the device) and many cosmic rays since the exposure time is 300 seconds. The example above produces this image:
Similarly, a flat can be generated by simulating an illumination pattern across the field. This can be done with this example that uses a flat SED illumination pattern.
phosim examples/flat
This catalog will close the dome and put on the dome light (telconfig 2). The dome light will have an intensity of 20.0 magnitudes per sq. arcsecond and a flat spectrum (domewav 0). If the dome wavelength is set to any other value other than 0, it simulates a monochromatic flat (i.e. laser illuminated). The commands are below.
vistime 15.0
telconfig 2
minsource 0
domeint 20.0
domewav 0
The flat image in this example is shown below